Routine Extractable Testing
From PQRI
Routine Extractables Testing should be performed on critical components using appropriate specifications and acceptance criteria. Analytical methods for Leachables Studies and Routine Extractables Testing should be fully validated according to accepted parameters and criteria.
Extraction and analytical methods for Routine Extractables Testing should be established based on methods developed in the Controlled Extraction Studies, and validated according to established parameters. Extractables profiles from these routine studies should be monitored for anomalous results. To aid in determination of anomalies, the sponsor should develop a profile specification, which should include acceptance criteria for known extractables as well as “unspecified” extractables, i.e., extractables not identified in qualitative Controlled Extraction Studies. Additionally, the sponsor should develop a procedure for investigating an obvious change in a component’s extractable profile which does not necessarily result in a batch failure (often termed an “out of trend investigation”). Following establishment of a correlation between leachables and extractables and a profile specification, Routine Extractables Testing for quality control should be performed. Based on Controlled Extraction Studies and leachables studies, acceptance criteria for leachables and extractables should be developed