Leachable - Extractable Correlations
Multiple
USP <1663>
are established when observed drug product leachables are linked both qualitatively and quantitatively to extractables from associated packaging/delivery systems, packaging components, or materials of construction
PQRI
The sponsor should establish a qualitative and quantitative correlation between extractables and leachables profiles. In establishing a correlation between profiles, the results of extraction studies on multiple batches of components and leachables studies on multiple batches of drug product over multiple stability storage time-points should be examined. To establish correlations, (i) leachables profiles from multiple (at least 3) drug product definitive registration batches (e.g., NDA stability batches, bio-batches, clinical batches, toxicology study batches) using specific batches of critical components, should be compared with qualitative and quantitative extractables profiles of those specific component batches, and (ii) leachables profiles from multiple drug product registration batches should be compared with extractables profiles from multiple batches of critical components (which may not have been used in the drug product registration batches). The results of leachables studies taken from multiple stability storage time-points and conditions should be correlated with results of extraction studies.
Extraction studies are conducted using multiple solvents/conditions so that asymptotic levels for extractables are achieved. Results from leachables studies should be obtained from samples incubated across the entire proposed shelf-life of the drug product, using appropriate ICH stability conditions. Extraction and leachable methods must be sufficiently sensitive to detect the full profile of extractables/leachables present above the AET, as well as be appropriately validated according to established parameters. Extractables profiles from quantitative studies should be compared with leachables profiles to determine extractables and leachables correlations. To establish a qualitative correlation between profiles, chemists must show that compounds detected in the leachables studies were also present in the Controlled Extraction Studies. To establish a quantitative correlation between profiles, chemists must show that levels of leachables obtained from leachables studies are generally less than the levels of extractables obtained from quantitative Controlled Extraction Studies.