Design Controls
From 21 CFR part 820.30(c)
(a) General. (1) Each manufacturer of any class III or class II device, and the class I devices listed in paragraph (a)(2) of this section, shall establish and maintain procedures to control the design of the device in order to ensure that specified design requirements are met.
(2) The following class I devices are subject to design controls:
(i) Devices automated with computer software; and
(ii) The devices listed in the following chart.
868.6810Catheter, Tracheobronchial Suction.878.4460Glove, Surgeon's.880.6760Restraint, Protective.892.5650System, Applicator, Radionuclide, Manual.892.5740Source, Radionuclide Teletherapy.